Development of an energy discriminate CR-39® nuclear track etch dosimeter for Radon-220 gas measurements
详细信息    查看全文
文摘
An energy discriminate CR-39® nuclear track etch dosimeter for use in a 220Rn and 222Rn gas monitor has been developed and experimentally assessed. It utilises a thin film of Mylar® C to attenuate the alpha particle energies to allow only the damage tracks created by the 8.785 MeV alpha particles emitted from 212Po of the 232Th decay chain to be registered in the CR-39® plaque, allowing for the direct measurement of 220Rn gas concentrations. The dosimeter was developed through a combination of experimental investigations and theoretical simulations using the Monte Carlo ion transport modelling program Stopping and Range of Ions in Materials (SRIM 2008). A film thickness of 54 μm has been shown to attenuate all alpha energies less then 7.7 MeV.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700