We propose a framework for the analysis of the repeatability and precision of corneal elevation and the repeatability of corneal thickness data measured by clinically used topographers and tomographers. The repeatability is given by the standard deviation of different measurements of the same eye for each data point. The differences between measurements can partially be explained by different positions of the eye at each measurement. The precision is given by the standard deviation of the data after correction for displacements (decentration, tilts, rotations). The applicability is demonstrated using measurements of the Pentacam HR [1] (Oculus Optikgeräte GmbH, Wetzlar, Germany) Scheimpflug tomographer. It provides anterior corneal elevation data with a precision that was almost limited by the axial resolution of 1 μm in the central 6 mm zone. This is approximately five times better than the precision of the posterior elevation. The results can be applied to surface approximation of the exported corneal elevation data and give a measure for the goodness of the fit.