Coronary artery calcium scoring: Influence of adaptive statistical iterative reconstruction using 64-MDCT
详细信息    查看全文
文摘

Objective

Assessment of coronary artery calcification is increasingly used for cardiovascular risk stratification. We evaluated the reliability of calcium-scoring results using a novel iterative reconstruction algorithm (ASIR) on a high-definition 64-slice CT scanner, as such data is lacking.

Methods and results

In 50 consecutive patients Agatston scores, calcium mass and volume score were assessed. Comparisons were performed between groups using filtered back projection (FBP) and 20-100 % ASIR algorithms. Calcium score was measured in the coronary arteries, signal and noise were measured in the aortic root and left ventricle. In comparison with FBP, use of 20 % , 40 % , 60 % , 80 % , and 100 % ASIR resulted in reduced image noise between groups (7.7 % , 18.8 % , 27.9 % , 39.86 % , and 48.56 % , respectively; p < 0.001) without difference in signal (p = 0.60). With ASIR algorithms Agatston coronary calcium scoring significantly decreased compared with FBP algorithms (837.3 ¡À 130.3; 802.2 ¡À 124.9, 771.5 ¡À 120.7; 744.7 ¡À 116.8, 724.5 ¡À 114.2, and 709.2 ¡À 112.3 for 0 % , 20 % , 40 % , 60 % , 80 % , and 100 % ASIR, respectively, p < 0.001). Volumetric score decreased in a similar manner (p < 0.001) while calcium mass remained unchanged. Mean effective radiation dose was 0.81 ¡À 0.08 mSv.

Conclusion

ASIR results in image noise reduction. However, ASIR image reconstruction techniques for HDCT scans decrease Agatston coronary calcium scores. Thus, one needs to be aware of significant changes of the scoring results caused by different reconstruction methods.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700