The study aims to address the prospective clinical utility of ultrasound compared with plain radiographs for assessing MCGR distractions.
This is a prospective study.
The study includes patients with early-onset scoliosis undergoing distractions after MCGR implant.
The distraction length on plain radiographs and ultrasound was measured.
This is a prospective study of patients treated with MCGR. Patients with both single- and dual-rod systems were included. Outpatient distractions were performed at monthly intervals, targeting 2 mm of distraction on each occasion. Assessment of distraction length was monitored by ultrasound at each visit; plain radiographs were taken every 6 months and were compared with ultrasound measurements.
Nine patients (5 female, 4 male), with a mean of 29 distractions (standard deviation [SD] ±14.3), were recruited. The mean distracted length per 6 months was 5.7 mm (SD ±3.6 mm) on plain radiographs and 5.2 mm (SD ±3.9 mm) on ultrasound for the concave rod, and 6.1 mm (SD ±3.6 mm) on plain radiographs and 5.9 mm (SD ±3.8 mm) on ultrasound for the convex rod. Excellent inter- and intra-rater reliabilities were observed for radiographic and ultrasound measurements. An excellent correlation was noted between the two imaging modalities (r=0.93; p<.0001).
This is the first prospective study to validate that ultrasound assessment of MCGR distraction lengths was highly comparable with that of plain radiographs. The present study has verified that ultrasound can be used to document length changes by distraction over time and that it had high clinical utility. Ultrasound can be a reliable alternative to plain radiographs, thereby avoiding radiation exposure and its potential detrimental sequelae in the developing child.