Tribocorrosion in shoulder arthroplasty humeral component retrievals
详细信息    查看全文
文摘
Tribocorrosion at the modular taper connections of total hip implants has been associated with trunnionosis and adverse local tissue reactions. Modularity is also widely used in shoulder arthroplasty implants, but little information exists about the potential for tribocorrosion. This study hypothesized that there would be mild or no tribocorrosion in a series of retrieved shoulder implants.

Methods

A total of 28 implants with a mean implantation time of 6.2 ± 6.0 years were evaluated using a validated damage scoring method. Implant tapers on the head and stem were divided into upper (deepest) and lower zones and independently scored for fretting and corrosion damage from 1 (none) to 4 (severe).

Results

Corrosion was present on 32% of heads and 38% of stems, whereas fretting was present on 36% of heads and 46% of stems. There was significantly greater (P = .02) corrosion in the lower zone of the retrieved stems (1.4 ± 0.5) than there was in the upper zone (1.1 ± 0.3). Correlation between the head and stem corrosion for lower zone was moderate (r = 0.41; P = .04).

Discussion

Tribocorrosion was present on the heads and stems of some of the retrieved shoulder implants examined in this study. The incidence of tribocorrosion in shoulder implants was lower than in reported cases of retrieved hip implants. The greatest damage was in the lower zone of the taper, where the connection may be exposed to the surrounding joint fluid. It remains to be seen whether this leads to any clinical presentation of trunnionosis.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700