Analytical evaluation of the novel Lumipulse G BRAHMS procalcitonin immunoassay
详细信息    查看全文
文摘
This study was designed to evaluate the analytical performance of the novel Lumipulse G1200 BRAHMS procalcitonin (PCT) immunoassay.Design and methodsThis analytical evaluation encompassed the calculation of the limit of blank (LOB), limit of detection (LOD), functional sensitivity, intra- and inter-assay imprecision, confirmation of linearity and a comparison with the Vidas BRAHMS PCT assay.ResultsThe LOB, LOD and functional sensitivity were 0.0010 ng/mL, 0.0016 ng/mL and 0.008 ng/mL, respectively. The total analytical imprecision was found to be 2.1% and the linearity was excellent (r=1.00) in the range of concentrations between 0.006–75.5 ng/mL. The correlation coefficient with Vidas BRAHMS PCT was 0.995 and the equation of the Passing and Bablok regression analysis was [Lumipulse G BRAHMS PCT]=0.76×[Vidas BRAHMS PCT]+0.04. The mean overall bias of Lumipulse G BRAHMS PCT versus Vidas BRAHMS PCT was −3.03 ng/mL (95% confidence interval [CI]: −4.32 to −1.74 ng/mL), whereas the mean bias in samples with PCT concentration between 0–10 ng/mL was −0.49 ng/mL (95% CI: −0.77 to −0.24 ng/mL). The diagnostic agreement was 100% at 0.5 ng/mL, 97% at 2.0 ng/mL and 95% at 10 ng/mL, respectively.ConclusionsThese results attest that Lumipulse G BRAHMS PCT exhibits excellent analytical performance, among the best of the methods currently available on the diagnostic market. However, the significant bias compared to the Vidas BRAHMS PCT suggests that the methods cannot be used interchangeably.