Three patients with PD (mean, 55 years) were scanned on 1.5-, 3-, and 7-T MRI before surgery. Tissue contrast and signal-to-noise ratio were measured. Target coordinates were noted on MRI and during surgery. Differences were analyzed with post-hoc analysis of variance.
The 7-T MRI demonstrated a significant improvement in tissue visualization (P < 0.005) and signal-to-noise ratio (P < 0.005). However, no difference in the target coordinates was found between the 7-T and the 3-T MRI.
Although the 7-T MRI enables a significant better visualization of the DBS target in patients with PD, we found no clinical benefit for the placement of the DBS leads.