Time-resolved and high-resolution MRA in a rabbit model of pulmonary embolism at 7 T: preliminary results
详细信息    查看全文
文摘

Purpose

Evaluate feasibility of using time-resolved and high-resolution, contrast-enhanced magnetic resonance angiography (MRA) at 7 T for characterization of an animal model of pulmonary embolism.

Methods

MRAs were performed in five rabbits using a 7-T MR scanner. Preceding the MR studies, each rabbit underwent a pulmonary artery catheterization with balloon placement. Two doses of gadodiamide were injected: first during a time-resolved MRA, immediately followed by a high-resolution acquisition. Balloon was then deflated, permitting reperfusion for 5 min. A second dose was then injected and another high-resolution MRA acquired. Measurements of signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR) and vessel cross-sections down to fourth-order branches were made, among other parameters.

Results

Occlusion was detected in all rabbits. Despite a TE of 0.58 ms for the time-resolved MRA, regions of nonuniform enhancement attributed to susceptibility effects at the 7-T field were observed in perfused lung. Mean SNR=7.5±3.3 and 134.2±46.5 for the lung and aorta, respectively, and mean CNR=126.7±46.4 for aorta versus lung were obtained. Diameters of vessels in lung that was never occluded were not statistically different from those in reperfused lung.

Conclusion

Results show that time-resolved and high-resolution MRA of the lung are feasible at 7 T and provide high SNR, CNR and resolution, but TEs smaller than 0.58 ms are required to avoid susceptibility artifacts in time-resolved MRAs.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700