This study sought to report the safety and performance data from the feasibility study with a novel direct annuloplasty system.
Seventy-one patients with moderate to severe FMR (mean 67.7 ± 11.3 years of age, left ventricular [LV] ejection fraction 34.0 ± 8.3%), on stable medical heart failure medication were prospectively enrolled.
Device success rate was 70.4% (n = 50 of 71). No intraprocedural death occurred. In patients receiving implants, 4 patients (8.9%) experienced cardiac tamponade. Thirty-day (n = 45) and 6-month (n = 41) rates for all-cause mortality, stroke, and myocardial infarction were 4.4%, 4.4%, and 0.0% and 12.2%, 4.9%, and 0%, respectively. At 6 months, nonurgent mitral surgery was performed in 1 patient (2.4%) and nonurgent percutaneous repair in 7 patients (17.1%). Echocardiographic core analysis after 6 months showed mitral regurgitation reduction in 50% of treated patients by a mean of 1.3 grades. Concerning mitral valve (MV) annular geometry, we found significant reduction of anterior-posterior (–0.31 ± 0.4 cm) and septal-lateral dimensions (–0.21 ± 0.3 cm), a decreased MV-tenting area (–0.57 ± 1.1 cm2) and increase in MV coaptation length (0.13 ± 0.2 cm). Transthoracic echocardiography indicated reverse LV remodeling with reduction of LV end-diastolic diameter (–0.20 ± 0.4 mm) and volume (–22 ± 39 ml). Treatment was associated with significant improvement in 6-min walking distances (56.5 ± 92.0 m) and improvements in New York Heart Association functional class III/IV at 6 months from 53.3% to 23.3%.
Percutaneous direct annuloplasty is feasible and safe in high-risk FMR patients. This treatment initiates LV reverse remodeling, and provides clinical improvement during 6 months after treatment. (Mitralign Percutaneous Annuloplasty First in Man Study; NCT01852149).