文摘
We have developed a simple, rapid and highly sensitive method for determining plasma concentrations of ganciclovir and/or acyclovir by using reversed-phase chromatography followed by pulsed amperometric detection. A linear relationship between the amount of ganciclovir (0.05–10 μg/ml plasma) or acyclovir (0.1–20 μg/ml plasma) and peak height ratio was obtained. The relative standard deviations of all standard curves were greater than or equal to 0.999. The limits of detection for ganciclovir and acyclovir quantitation were 10 ng/ml and 50 ng/ml (signal/noise >3), respectively. Daily fluctuations of plasma standard curves (n