A sub-two minutes method for monoclonal antibody-aggregate quantification using parallel interlaced size exclusion high performance liquid chromatography
详细信息    查看全文
文摘
In process development and during commercial production of monoclonal antibodies (mAb) the monitoring of aggregate levels is obligatory. The standard assay for mAb aggregate quantification is based on size exclusion chromatography (SEC) performed on a HPLC system. Advantages hereof are high precision and simplicity, however, standard SEC methodology is very time consuming. With an average throughput of usually two samples per hour, it neither fits to high throughput process development (HTPD), nor is it applicable for purification process monitoring. We present a comparison of three different SEC columns for mAb-aggregate quantification addressing throughput, resolution, and reproducibility. A short column (150 mm) with sub-two micron particles was shown to generate high resolution (?.5) and precision (coefficient of variation (cv) < 1) with an assay time below 6 min. This column type was then used to combine interlaced sample injections with parallelization of two columns aiming for an absolute minimal assay time. By doing so, both lag times before and after the peaks of interest were successfully eliminated resulting in an assay time below 2 min. It was demonstrated that determined aggregate levels and precision of the throughput optimized SEC assay were equal to those of a single injection based assay. Hence, the presented methodology of parallel interlaced SEC (PI-SEC) represents a valuable tool addressing HTPD and process monitoring.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700