Central nervous system MRI and cardiac implantable electronic devices
详细信息    查看全文
文摘
As the population ages and indications for MRI increase, it is estimated that 50 to 75% of patients with a cardiac implantable electronic device (CIED) – pacemaker (PM) or implanted cardiac defibrillator (ICD) – will need an MRI during their CIED's lifetime. Three categories of materials are defined: MRI compatible, MRI non-compatible, and MRI conditional. MRI compatible CIEDs without electrodes have been developed, but do not allow battery changes, so that they are exclusively indicated for patients whose life expectancy is less than that of the battery (6–7 years). For MRI conditional CIEDs, all manufacturers publish restrictions. These restrictions can relate to the patient (size, position in the MRI, body temperature), the MRI parameters (magnetic field), or the examination in itself (gradients, specific absorption rate, duration, isocenter). The neuroradiologist can expect to be confronted with the issue of MRI in patients with a CIED. The purpose of this review is to provide them with updated information on MRI and CIEDs.