Improvement on AMS measurement method for 182Hf
详细信息    查看全文
文摘
In order to improve the detection efficiency and sensitivity for 182Hf measurement in HI-13 AMS systems at China Institute of Atomic Energy (CIAE), a new method was developed. Main features of the method include: (1) the installation of a new injection system with high mass resolution, (2) the establishment of a method based on solid-phase reaction to prepare HfF4 samples from HfO2. The experimental results show that the, mass resolution (M/ΔM) of the new injector system can reach nearly 700; the beam current of 180HfF5- ions from HfF4 samples produced by solid-phase reaction is about two times higher than that of previous method.