Comparative analysis for measurement precision of phase retrieval wavefront sensor
详细信息    查看全文
文摘
In order to analyze the performance of phase retrieval wavefront sensor (PRWS), an experimental based on the PRWS method has been built. In this paper, we utilize liquid crystal space light modulator (LC-SLM) to produce single aberration and random aberration, and validate the ability of PRWS measurement for any aberrations. Mathematic model of phase retrieval measurement is constructed based on the diffraction theory, and solved using gradient based optimization algorithm. Experimental results demonstrate that good agreement is obtained among the errors distribution, PV value and RMS value of ZYGO interferometer. The measurement precision of wavefront is 3/1000 wavelength RMS. Therefore, the feasibility and accuracy of the proposed method can be confirmed.