Paired data were collected on 22 patients (aged 67.5 ¡À 8.3 years, 16 male) with refractory heart failure, NYHA class III/IV symptoms, sinus rhythm, LBBB and a broad QRS complex >120 ms. All patients underwent implantation of a biventricular pacemaker and were randomised to eight weeks of factory pacing mode (Mode 1) or echocardiographically-guided pacing mode (Mode 2), followed by eight weeks in the alternate mode, in a randomised blinded crossover design.
Peak oxygen consumption, 6 min walk distance, NYHA class and quality of life scores improved after biventricular pacing, but no significant difference was found between the two modes, with the exception of peak oxygen consumption score (baseline: 14.8 ¡À 0.9, Mode 1: 14.6 ¡À 1.2, Mode 2: 16.1 ¡À 1.2 mL/kg/min), which was better in Mode 2 than Mode 1 (p 0.003).
Transthoracic echocardiographic optimisation of AV and VV delays following biventricular pacing may offer additional clinical benefit in an unselected group of patients when compared with factory settings.