The values of the cardiac markers assessed by the respective assays correlated (cTnT/cTnI Rho = 0.94, myoglobin Rho = 0.87, CK-MB Rho = 0.84). If values were dichotomised, we found a high concordance of test positive and negative classified patients by troponins with the respective assays.
There is strong evidence that the Innotrac Aio!™ system for cTnI measurement can be used reliably.