Comparative evaluation of the QUANTIPLEX HIV-1 RNA 2.0 and 3.0 (bDNA) assays and the AMPLICOR HIV-1 MONITOR v1.5 test for the quantitation of human immunodeficiency virus type 1 RNA in plasma
详细信息    查看全文
文摘
HIV-1 RNA measurements from 84 plasma specimens obtained with the QUANTIPLEX HIV-1 RNA 2.0 and 3.0 (bDNA) assays (Chiron Diagnostics, Emeryville, CA) and with the AMPLICOR HIV-1 MONITOR™ Test, version 1.5 with ultra-sensitive specimen preparation (Roche Diagnostic Systems, Inc., Branchburg, NJ) were compared. The absolute RNA values of tested specimens differed significantly between bDNA 2.0 and bDNA 3.0 or Monitor v1.5 measurements (Wilcoxon signed-rank test P<0.001). Results generated with bDNA 3.0 or with Monitor v1.5 were approximately twofold greater than those generated with bDNA 2.0, with smaller differences at higher HIV-1 RNA levels and greater differences at RNA levels below 1000 copies per ml. Although highly correlated (r