Relative dose and vascular response after drug-eluting stent implantation: A dosimetric 3D-intravascular ultrasound study
详细信息    查看全文
文摘
In drug-eluting stents (DESs), the theoretical drug dose exposed to the vessel wall per stent surface area may vary due to the fixed loading dose and differences in the stent surface area once expanded in varying vessel sizes. The aim of this study was to evaluate the potential effects of different dose intensities, as estimated by 3D-IVUS dosimetry, on vascular response after DES implantation.

Methods

Follow-up (6–9 months) 3D-IVUS was performed in 840 coronary lesions treated with a single DES of the following types: sirolimus (SES, n = 148), paclitaxel (PES, n = 162), Endeavor zotarolimus (E-ZES, n = 233), Resolute zotarolimus (R-ZES, n = 147), and everolimus (EES, n = 150). Volume index (volume/length, mm3/mm) was obtained for vessel, lumen, plaque, stent, and neointima. In each lesion, exposed dose intensity was calculated as known loading dose divided by measured luminal surface area of the stented segment. Lesions were divided into tertiles based on the exposed dose intensity: high, medium, and low dose groups.

Results

The exposed dose intensity ranged 0.74–1.76 μg/mm2 for SES, 0.41–1.18 μg/mm2 for PES, 0.71–1.57 μg/mm2 for E-ZES, 0.72–1.63 μg/mm2 for R-ZES, and 0.40–0.99 μg/mm2 for EES. All types of DES showed no significant difference in neointimal hyperplasia among the 3 groups, except that E-ZES showed significantly less neointimal hyperplasia in the high dose group.

Conclusions

Detailed 3D-IVUS revealed significant lesion-to-lesion variability in dose intensity exposed to the vessel wall following DES implantation. However, the major types of DES appear to yield equally effective neointimal suppression, despite the varying dose intensity, except for E-ZES.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700