This study was comprised of 160 patients with AF (54% paroxysmal, mean: 24.0±2.9 kg/m2). Eighty patients underwent ablation accompanied by ETM. The primary endpoint was defined as the occurrence of ETI assessed by endoscopy within 5 d after the AF ablation. The secondary endpoint was defined as AF recurrence after a single procedure. If the esophageal temperature probe registered >39 °C, the radiofrequency (RF) application was stopped immediately. RF applications could be performed in a point-by-point manner for a maximum of 20 s and 20 W. ETI was defined as any injury that resulted from AF ablation, including esophageal injury or periesophageal nerve injury (peri-ENI).
The incidence of esophageal injury was significantly lower in patients whose AF ablation included ETM compared with patients without ETM (0 [0%] vs. 6 [7.5%], p=0.028), but not the incidence of peri-ENI (2 [2.5%] vs. 3 [3.8%], p=1.0). AF recurrence 12 months after the procedure was similar between the groups (20 [25%] in the ETM group vs. 19 [24%] in the non-ETM group, p=1.00).
Catheter ablation using ETM may reduce the incidence of esophageal injury without increasing the incidence of AF recurrence but not the incidence of peri-ENI.