详细信息    查看全文
文摘
To evaluate changes in the aqueous humor temperature using 2 phacoemulsification units (Stellaris 28.5 kHz device and Whitestar Signature 40 kHz device).

Setting

Nippon Medical School, Musashikosugi Hospital, Kawasaki City, Kanagawa, Japan.

Design

Experimental study.

Methods

Aqueous humor temperatures were measured with a temperature probe set in the anterior chamber during ultrasound (US) oscillation in porcine eyes under 5 conditions.

Results

Continuous longitudinal oscillation caused a rapid rise in aqueous humor temperature, while the pulse and elliptical modes suppressed temperature elevation. Reducing the number of US tip vibrations did not reduce the temperature in the anterior chamber. However, raising the vacuum setting allowed the aspirations to rise to the set value, thereby lowering the temperature in the anterior chamber.

Conclusion

Because differences in the phacoemulsification settings can lead to temperature elevations in the anterior chamber, surgeons must carefully monitor these settings to prevent corneal tissue damage.

Financial Disclosure

No author has a financial or proprietary interest in any material or method mentioned.