Effect of pH on the stability of kidney injury molecule 1 (KIM-1) and on the accuracy of its measurement in human urine
详细信息    查看全文
文摘

Background

Urinary KIM-1 is a novel biomarker for tubular kidney damage, however little is known about its stability. The goal of this study is to examine the effect of urinary pH on the stability of KIM-1.

Methods

Urine samples were collected from 45 volunteers. Samples were aliquoted, adapted to different pH values (range 4 to 9) and stored at −80 °C. After thawing, each aliquot was divided into two, of which one was used to measure KIM-1 (human tim-1/kim-1/Havcr Elisa kit; R&D systems) at the same pH at which it was stored, while the other was readapted to pH 7 before measurement.

Results

KIM-1 values of aliquots of the same sample are stable when stored at pH 6, 7 and 8 whereas at lower and higher storage pH, KIM-1 levels decrease significantly. When samples are readjusted to a neutral pH just before KIM-1 measurement, there are no longer significant differences between KIM-1 in aliquots stored at different pH values.

Conclusions

No effect of urinary pH on the stability of KIM-1 was seen. However, the only commercially available human tim-1/kim-1/Havcr Elisa kit of RD systems is pH dependent and we therefore suggest samples should be adjusted to neutral pH before measurement.