An innovated method based on DANP-IMS was put forward.
Influence of moisture was effectively reduced via an on-line semiconductor cooling inlet.
LOQs for humid H2S were achieved to ≤79.0 ng L−1.
Sewer H2S content in manholes was monitored in real time and its correlation with the resident daily activities was observed.