Comparison of Tibial Insert Polyethylene Damage in Rotating Hinge and Highly Constrained Total Knee Arthroplasty: A Retrieval Analysis
详细信息    查看全文
文摘
This study compared the damage scores and damage patterns in 19 tibial inserts from rotating hinge (RH) implants with 19 inserts from highly constrained (HC) implants. Each insert was divided into 16 damage zones and each zone was subjectively graded from a scale of 0–3 for seven different damage modes. The overall damage scores were comparable for the two groups (RH: 64.1 ± 15.4; HC: 66.1 ± 29.0; P = 0.59). The HC group, however, had greater post damage (compared to the post-hole of RH) while the RH group had greater backside damage. The pattern of damage was also different, with burnishing and cold flow being more common in HC group while pitting, scratching and embedded debris were more common in the RH group.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700