Trunnionosis: Does Head Size Affect Fretting and Corrosion in Total Hip Arthroplasty?
详细信息    查看全文
文摘
Wear and tribocorrosion at the modular head–neck taper interface may be a cause of failure in metal-on-polyethylene total hip arthroplasty (THA). The present investigation endeavored to elucidate the effect of femoral head diameter on fretting and corrosion in retrieved head–neck tapers.

Methods

A retrieval analysis of THA prostheses in vivo for a minimum of 1 year was performed. Twenty-three femoral heads of 32-mm diameter were matched with 28-mm heads based on time in vivo and head length (−3 mm to +8 mm). All included implants featured a single taper design from a single manufacturer. Fretting and corrosion damage scoring was performed for each implant under stereomicroscopic visualization.

Results

Head diameter was observed to affect fretting (P = .01), with 32-mm femoral heads exhibiting greater total fretting scores than 28-mm heads. Fretting damage was greatest (P = .01) in the central concentric zone of the femoral head bore tapers, regardless of head diameter, length, or stem offset. No significant effect on total corrosion scores was observed for any head or stem variable. Retrieved implant total corrosion scores were positively correlated (ρ = 0.51, P < .001) with implantation time.

Conclusion

Increased femoral head diameter in THA may produce greater fretting damage owing to and increased head–neck moment arm. There is no associated increase in corrosion with 28-mm and 32-mm heads of this taper design. The longer a THA prosthesis is implanted, the greater the risk of damage due to corrosion.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700