Status of mass spectrometric radiocarbon detection at ETHZ
详细信息    查看全文
文摘
A prototype of a mass spectrometric radiocarbon detection instrument without accelerator stage was built for the first time and set into operation at ETH Zurich. The system is designed as an experimental platform to optimize performance of 14C detection at low ion energies and to study the most relevant processes that may limit system performance. The optimized stripper unit incorporates differential pumping to maintain a low gas outflow and a revised tube design to better match the phase space volume of the ion beam at low energies. The system is fully operational and has demonstrated true radiocarbon dating capabilities. The overall beam transmission through the stripper tube is about 40% for the 1+ charge state. Radiocarbon analyses with an overall precision of 0.6% were obtained on a single sample under regular measurement conditions. By analyzing multiple targets of the same sample material an uncertainty level of 0.3% has been reached. The background level corresponds to a radiocarbon age of 40,000 years.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700