The efficacy and safety of epidural filter catheters when used for post-operative intra-articular anaesthesia
详细信息    查看全文
文摘

Epidural filters prevent bacterial transit during high volume multiple infusions.

Manual controlled administration (20 ml/min) generated no filter ruptures.

Automated syringe driver administration generated no filter ruptures.

Manual forced administration generated 93% filter rupture at ~ 620 kPa.

We recommend using automated devices for post-operative intra-articular infusions.