Clinical Validation of a Novel Speckle-Tracking¨CBased Ejection Fraction Assessment Method
详细信息    查看全文
文摘
| Figures/TablesFigures/Tables | ReferencesReferences

Background

The aim of this study was to determine the feasibility, accuracy, and reproducibility of a novel tracking-based echocardiographic ejection fraction (EF) assessment method in comparison with traditional methods based on magnetic resonance imaging and echocardiography.

Methods

In a prospective assessment, apical echocardiographic grayscale image loops from 81 patients were read in random order by four experienced readers, blinded to any data of the cases. In three separate sessions, EFs were estimated using biplane tracking-based assessment and according to the modified Simpson¡¯s rule, as well as by visual interpretation in three apical views. Data were compared with a reference EF derived from echocardiography and magnetic resonance imaging.

Results

On average, no significant difference was found between EF estimates of the different methods. Tracking-based EF assessments were possible in 90 % of the patients. Tracking-based EF assessments showed slightly higher deviations from the reference EF than the modified Simpson¡¯s rule, while interobserver and intraobserver variability of tracking-based assessment were significantly better. Visual interpretation allowed the fastest EF assessment. Tracking-based EF assessment was approximately twice as fast as the modified Simpson¡¯s rule.

Conclusions

Tracking-based EF assessment is feasible, has lower interobserver and intraobserver variability, and is faster than traditional echocardiographic EF quantification. Its minimal demand on user interaction makes it a favorable alternative to traditional echocardiographic approaches, with a particular clinical advantage when reliable follow-up measurements are needed.