A calibration setting with uncertainty measurements for passive/active radon monitors using flow-through source type
详细信息    查看全文
文摘
Based upon the steady flow method, a setting has been introduced for fast calibration of almost all types of active and passive radon detectors. In this method, a traceable radon reference flow-through sources is connected to a portable rectangular cubic transparent chamber (volume of 50 L) which is completely sealed and has 6 valves on its sides. The initializing calibration times are less than 150 min by this chamber. The radon concentration in the chamber is in consistent with the source activity. The Best Measurement Capability (BMC) of the method has been calculated less than 5 % at 68 % confidence level. As well by using grab-sampling method, the difference of radon concentration of different points in the chamber was measured less than 1.2 % . So the flow-through sources can be developed for a fast calibration of almost all types of the radon monitors with an acceptable uncertainty using this calibration setting.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700