文摘
We have studied the neutron response of a scintillation detector consisting of a 14cm long, hexagonal-shaped BaF2-crystal with an inner diameter of 8.75cm coupled to an EMI9821QB photomultiplier tube. The detector was exposed to calibrated quasi-monoenergetic neutron fields obtained from 7Li(p,n)7Be reactions. The measurements were performed at neutron energies of 45, 60, 96, 147 and 198MeV as given by the energies of the incident protons. The experimental pulse-height spectra of the BaF2-detector are compared with Monte Carlo simulations using the FLUKA code. The detection efficiency of the BaF2-detector in the energy range of 45–198MeV was determined as a function of the discriminator threshold and compared to the literature data. At neutron energies above 100MeV the detection efficiency of the BaF2-detector was found to be a factor of two higher than that of an NE213-detector of comparable size.