Selection and fabrication of a non-woven polycarbonate urethane cover for a tissue engineered airway stent
详细信息    查看全文
文摘
One of the major problems in end-stage bronchotracheal cancer is stenosis of the upper airways, either due to luminal ingrowth of the tumor or mucus plugging. Airway stents that suppress tumor ingrowth and sustain mucociliary transport can alleviate these problems in end-stage bronchial cancer. We evaluated different types of polymeric covers for a tissue engineered airway stent. The distinguishing feature of this stent concept is that respiratory epithelial cells can grow on the luminal surface of the stent which facilitates mucociliary clearance. To facilitate growth of epithelial cells at the air-liquid interface of the stent, we developed a polyurethane cover that allows transport of nutrients to the cells. Nonwoven polycarbonate urethane (PCU) covers were prepared by a spraying process and evaluated for their porosity and glucose permeability. Respiratory epithelial cells harvested from sheep trachea were cultured onto the selected PCU cover and remained viable at the air-liquid interface when cultured for 21 days. Lastly, we evaluated the radial force of a PCU-covered nitinol stent, and showed the PCU covers did not adversely affect the mechanical properties of the stents for their intended application in the smaller bronchi. These in vitro data corroborate the design of a novel airway stent for palliative treatment of bronchotracheal stenosis by combination of stent-technology with tissue-engineered epithelial cells.