文摘
A novel glassy carbon electrode (GCE) modified with a film of ferroferric oxide nanoparticles (Fe3O4 NPs) over a film of gold nanoparticles (Au NPs), both stabilized in a polymer solution of 3-n-propyl-4-picolinium silsesquioxane chloride (Si4Pic+Cl−), is presented for the electroanalytical determination of bisphenol A (BPA). At the modified electrode, cyclic voltammograms of BPA in B-R buffer (pH 9.0) showed a diffusion-controlled irreversible oxidation peak at around +0.470 V. Under optimized experimental conditions for differential pulse voltammetry experiments, the peak current was found to vary linearly with the concentration of BPA in the range of 20–1400 nmol L−1 (R2 = 0.996) with a detection limit of 7.0 nmol L−1. The modified electrode Fe3O4 NPs-Si4Pic+Cl−/Au NPs-Si4Pic+Cl−/GCE was successfully employed for the determination of BPA in different commercial plastic samples. Satisfactory recoveries were obtained in the range of 90–120%, in agreement with the comparative technique (UV–vis spectroscopy).