The applicability of correction factor for corneal thickness on non-contact tonometer measured intraocular pressure in LASIK treated eyes
详细信息    查看全文
文摘
To determine the applicability of central corneal thickness (CCT) based correction factor for non-contact tonometer (NCT) measured intraocular pressure (IOP) readings.

d="absSec_2">Method

d="sp0010">A prospective, non-randomized study involved 346 eyes of 173 consecutive patients with age ⩾21 years undergoing laser-assisted in situ keratomileusis (LASIK) for myopia and/or myopic astigmatism. The CCT and IOP were measured before and after the LASIK procedure. The IOP pre and post-LASIK was compared after applying the correction factor for CCT. Patients not completing the 3 month postoperative follow-up were excluded.

d="absSec_3">Results

d="sp0015">The median spherical equivalent before undergoing LASIK was −4.25D (inter-quartile range, −3.25D). The mean preoperative CCT was 536.82 ± 33.71 μm which reduced to 477.55 ± 39.3 μm (p < 0.01) post-LASIK. The mean IOP reduced from a preoperative value of 14.6 ± 2.32 mmHg to 10.64 ± 2.45 mmHg postoperatively (p < 0.01). On applying correction for the corneal thickness, the pre and postoperative IOP was 15.14 ± 2.8 mmHg and 15.37 ± 2.65 mmHg (p = 0.06) respectively with a strong positive correlation (r = 0.7, p < 0.01). Three hundred eyes (86.7%) had an absolute difference in IOP of less than 3.0 mmHg post-CCT correction which is within the retest variability of NCT. Only 46 eyes (13.3%) had an absolute difference of more than 3.0 mmHg.

d="absSec_4">Conclusion

d="sp0020">The modified Ehler’s correction algorithm used in this study can be effectively applied in the normal IOP range in a majority of patients.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700