Sulcus anatomy and diameter in pseudophakic eyes and correlation with biometric data: Evaluation with a 50 MHz ultrasound biomicroscope
详细信息    查看全文
文摘
| Figures/TablesFigures/Tables | ReferencesReferences

Purpose

To evaluate the sulcus anatomy and possible correlations between sulcus diameter and white-to-white (WTW) diameter in pseudophakic eyes, data that may be important in the stability of add-on intraocular lenses (IOLs).

Setting

University Eye Hospital, Tuebingen, Germany.

Design

Case series.

Methods

In pseudophakic eyes, the axial length (AL) and horizontal WTW were measured by the IOLMaster device. Cross-sectional images were obtained with a 50 MHz ultrasound biomicroscope on the 4 meridians: vertical, horizontal (180 degrees), temporal oblique, and nasal oblique. Sulcus-to-sulcus (STS), angle-to-angle (ATA), and sclera-to-sclera (ScTSc) diameters were measured. The IOL optic diameter (6.0 mm) served as a control. To test reliability, optic measurements were repeated 5 times in a subset of eyes.

Results

The vertical ATA and STS diameters were statistically significantly larger than the horizontal diameter (P=.0328 and P=.0216, respectively). There was no statistically significant difference in ScTSc diameters. A weak correlation was found between WTW and horizontal ATA (r = 0.5766, P<.0001) and between WTW and horizontal STS (r = 0.5040, P=.0002). No correlation was found between WTW and horizontal ScTSc (r = 0.2217, P=.1217).

Conclusions

The sulcus anatomy had a vertical oval shape with the vertical meridian being the largest, but it also had variation in the direction of the largest meridian. The WTW measurements showed a weak correlation with STS. In pseudophakic eyes, Soemmerring ring or a bulky haptic may affect the ciliary sulcus anatomy.

Financial Disclosure

No author has a financial or proprietary interest in any material or method mentioned.