Experimental study on stability of different solid contact arrangements of ion-selective electrodes
详细信息    查看全文
文摘
Comparison of potentials stability of different types of solid contact lead selective electrodes is presented. Conducting polymer based sensors (hydrophilic and conducting – poly(3,4-ethylenedioxythiophene) or hydrophobic and semiconducting-polyoctylthiophene) were studied in parallel with coated wire and hydrogel (poly(hydroxyethylmethacrylate)) contact electrodes. The within day and between days potential stability was compared, highlighting the effect of sensor storing conditions. The obtained results clearly demonstrate that different conclusions on stability of sensors’ potential can be drawn depending on experimental protocol applied. Polyoctylthiophene based contacts show superior within day stability with no influence of dry storage. On the other hand, a chronopotentiometric method of stability evaluation clearly prefers poly(3,4-ethylenedioxythiophene) based sensors, pointing to smallest resistance and polarizability. It is clearly shown that the choice of experimental conditions applied to test stability can favor particular type of contact used.

The inductively coupled plasma mass spectrometry with laser ablation (LA-ICP-MS) experiments conducted for different arrangements tested has shown that for poly(3,4-ethylenedioxythiophene) and poly(hydroxyethylmethacrylate) type contacts longer contact time with lead(II) solution results in changes in the elemental composition of the transducer layer. On the other hand, in line with high stability observed under potentiometric conditions, no changes were seen for polyoctylthiophene based transducer.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700