Multipolar radiofrequency ablation using internally cooled electrodes in ex vivo bovine liver: Correlation between volume of coagulation and amount of applied energy
详细信息    查看全文
文摘

Purpose

To evaluate the relationship between applied energy and volume of coagulation induced by multipolar radiofrequency (RF) ablation.

Methods and materials

Multipolar RF ablations (n = 80) were performed in ex vivo bovine liver. Three bipolar applicators with two electrodes located on each applicator shaft were placed in a triangular array. The power-output (75-225 W) and the distance between the different applicators (2, 3, 4, 5 cm) were systematically varied. The volume of confluent white coagulation and the amount of applied energy were assessed. Based on our experimental data the relationship between the volume of coagulation and applied energy was assessed by nonlinear regression analysis. The variability explained by the model was determined by the parameter r2.

Results

The volume of coagulation increases with higher amounts of applied energy. The maximum amount of energy was applied at a power-output of 75 W and an applicator distance of 5 cm. The corresponding maximum volume of coagulation was 324 cm3 and required an application of 453 kJ. The relationship between amount of applied energy (E) and volume (V) of coagulation can be described by the function, V = 4.39E0.7 (r2 = 0.88). By approximation the volume of coagulation can be calculated by the linear function V = 0.61E + 40.7 (r2 = 0.87).

Conclusion

Ex vivo the relationship between volume of coagulation and amount of applied energy can be described by mathematical modeling. The amount of applied energy correlates to the volume of coagulation and may be a useful parameter to monitor multipolar RF ablation.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700