The 12 lead ECG was measured in 18 healthy and semi-recumbent subjects using the standard and Takuma modified limb placements.
In the frontal plane we demonstrate that the modification of limb electrode placement produces small Q, R and T wave amplitude and QRS axis changes that are statistically but not clinically significant. In the transverse plane it produces no statistically or clinically significant changes in the ECG or in ST segment morphology, P¨CR, QRS or QTc intervals.
We provide better and more robust evidence that routine modification of limb electrode placement produces only minor changes to the ECG waveform in healthy subjects. These are not clinically significant according to the 2009 guidelines and thus have no effect on the clinical specificity of the 12 lead ECG.