Outcomes of the Recalled Articular Surface Replacement Metal-on-Metal Hip Implant System: A Systematic Review
详细信息    查看全文
文摘
The aim of this review was to synthesize the main findings of clinical studies that have evaluated outcomes of the articular surface replacement (ASR) Hip System.MethodsWe performed a systematic literature search to identify all articles published between January 2008 and June 2015 that included ASR hip resurfacing arthroplasty (ASR HRA) or ASR total hip arthroplasty (ASR XL THA) outcomes according to the PRISMA statement.ResultsA total of 56 studies were assessed. The prevalence of adverse local tissue reactions (ALTRs) and revision rates were found to be high. ALTR prevalence varied from 12.5% to 69% (mean, 33.5%). Mean revision rate for any reason at 4-year to 7-year follow-up was 13.8% (range, 5.6%-31%) for ASR HRA and 14.5% (range, 0%-37%) for ASR XL THA. Femoral head size <53 mm was found to correlate with higher blood metal ion levels. Femoral head size >44 mm was not associated with higher ALTR prevalence or revision rates in ASR XL THA. High blood metal ion levels (>7 μg/L Co, >7 μg/L Cr) were associated with higher failure rates and bearing-related complications. The role of cup positioning was found to be controversial.ConclusionALTR prevalence and failure rates were high. High blood metal ion levels were a risk factor for ALTR and failure. Surprisingly, the role of cup positioning and large femoral head size in ASR XL THA were controversial. These findings should be considered in the clinical follow-up and risk stratification of patients with the ASR Hip System.