Combined Use of a Patient Dose Monitoring System and a Real-Time Occupational Dose Monitoring System for Fluoroscopically Guided Interventions
详细信息    查看全文
文摘
To determine the effect on patient radiation exposure of the combined use of a patient dose monitoring system and real-time occupational dose monitoring during fluoroscopically guided interventions (FGIs).

Materials and Methods

Patient radiation exposure, in terms of the kerma area product (KAP; Gy ∙ cm2), was measured in period 1 with a patient dose monitoring system, and a real-time occupational dose monitoring system was additionally applied in period 2. Mean/median KAP in 19 different types of FGIs was analyzed in both periods for two experienced interventional radiologists combined as well as individually. Patient dose and occupational dose were correlated, applying Pearson and Spearman correlation coefficients.

Results

Although FGIs were similar in numbers and types over both periods, a substantial decrease was found for period 2 in total mean ± SD/median KAP for both operators together (period 1, 47 Gy ∙ cm2 ± 67/41 Gy ∙ cm2; period 2, 37 Gy ∙ cm2 ± 69/34 Gy ∙ cm2) as well as for each individual operator (for all, P < .05). Overall, KAP declined considerably in 15 of 19 types of FGIs in period 2. Mean accumulated dose per intervention was 4.6 µSv, and mean dose rate was 0.24 mSv/h. There was a strong positive correlation between patient and occupational dose (r = 0.88).

Conclusions

Combined use of a patient dose monitoring system and a real-time occupational dose monitoring system in FGIs significantly lessens patient and operator doses.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700