In this study, the performance of two commercial real-time PCR kits in Beijing was evaluated. Analysis of clinical samples showed that the positive detection rate for the AgPath-ID? kit (38.2 % ) was significantly higher than that for the Da An H1N1 kit (30.0 % ) (McNemar's chi-square test, P = 0.000). The limit of detection (LOD) of the AgPath-ID? kit was 102, 102, and 103 copies/reaction for the Influenza A (set 1), H1N1 Influenza A (set 2) and H1N1 Influenza A Sub H1 (set 3) genes, respectively, whereas the LOD of the Da An kit was 103 copies/reaction for both H1 and N1 genes. Although the AgPath-ID? kit exhibited a significantly higher detection rate for pH1N1 than the Da An kit, cross-reactivity to A/PR8/34 was found for the AgPath-ID? kit for H1N1 Influenza A (set 2).