Pressure behavior of different PEEK materials for dental implants
详细信息    查看全文
文摘
Due to its mechanical properties, the biocompatible high-performance material PEEK (polyetheretherketone) and PEEK-based compounds may represent viable alternatives to titanium in the field of dental implantology. Therefore we performed static pressure tests with 11 PEEK materials (two unfilled grades, two grades filled with titanium-dioxide-powder, two grades filled with barium-sulfate-powder, two grades reinforced with short carbon fibers, one grade reinforced with glass fibers and two grades reinforced with continuous carbon fibers) in the form of cylindrical specimens with a diameter of 4, 5 and 6 mm. The specimens had a height to diameter ratio of 2:1 and were therefore 8, 10 and 12 mm high. The parameters elastic modulus, elastic limit and pressure strength were evaluated. The elastic moduli ranged between 2.65±0.03 GPa for specimens of a titanium-dioxide-filled grade and 106.71±14.83 GPa for specimens reinforced with continuous carbon fibers. The elastic limits ranged between 808.1±42.44 N for specimens of a barium-sulfate-filled grade and 7256.4±519.86 N for specimens reinforced with continuous carbon fibers. The lowest pressure strength of 122.77 MPa was observed for specimens of an unfilled grade, whereas the highest pressure strength of 712.67±66.02 MPa could be evaluated for specimens containing continuous carbon fibers. Regarding the maximum bite force of a first molar, all tested materials seem to be suitable for the use as dental implants.