Improving the Definition of Tumor Bed Boost With the Use of Surgical Clips and Image Registration in Breast Cancer Patients
详细信息    查看全文
文摘

Purpose

To evaluate the accuracy of a boost technique.

Methods and Materials

Twenty-two patients underwent tumorectomy with placement of two or more clips in the surgical cavity before breast remodeling. Preoperative and postoperative computed tomography scans, with match-point registration, were performed on all patients. The relationship between the location of the gross tumor volume (GTV), defined on the preoperative scan, and clip clinical target volume (CTV) (clips with a 5-mm margin on the postoperative scan) was then studied, by use of commercial volume analysis software.

Results

Of the patients, 4 had two clips, 2 had three clips, 8 had four clips, and 8 had five clips. The median GTV was 1.06 mL (range, 0.2–5.3 mL); clip CTV ranged from 2.4 to 21.5 mL. Volumetric analysis showed that in 7 cases (32 % ), there was no intersection between the GTV and the clip CTV, with the following distribution: 4 patients with two clips, 1 patient with three clips, 1 patient with four clips, and 1 patient with five clips. The common contoured volume was defined as the percent ratio between the intersection of the GTV and clip CTV and the GTV. It was found to be significantly increased if three or more clips were used vs. only two clips (common contoured volume, 35.45 % vs. 0.73 % ; p = 0.028). Finally, the GTV and clip CTV volume relationship can be presented as follows: 12.5 % to 33 % overlap in 8 patients (36.4 % ), 50 % to 75 % in 5 patients (22.7 % ), and greater than 90 % in 2 patients (9 % ).

Conclusions

The use of three or more clips during tumorectomy increases the accuracy of tumor bed delineation.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700