spara0015">Participants performed a range of activities (sitting, lying, walking, standing) in both a controlled and a home setting while wearing the accelerometer. Activities in the controlled setting were performed in a scripted sequence. Activities in the home setting were performed in an unscripted manner. Analyzed accelerometer data were compared against video observation as the reference measure.
spara0020a">Independent-living and long-term-care retirement village.
spara0020">Older people (N=22; mean age ± SD, 88.1±5y) residing in long-term-care and independent-living retirement facilities.
spara0025">Not applicable.
spara0030">The level of agreement between video observation and the accelerometer for the total duration of each activity, and second-by-second correspondence between video observation and the accelerometer for each activity.
spara0035">The median absolute percentage errors between video observation and the accelerometer were <1% for locomotion and lying. The absolute percentage errors were higher for sitting (median, −22.3%; interquartile range [IQR], −62.8% to 10.7%) and standing (median, 24.7%; IQR, −7.3% to 39.6%). A second-by-second analysis between video observation and the accelerometer found an overall agreement of ≥85% for all activities except standing (median, 56.1%; IQR, 34.8%–81.2%).
spara0040">This single-device accelerometer provides a valid measure of lying and locomotion in people aged >80 years. There is an error of approximately 25% when discriminating sitting from standing postures, which needs to be taken into account when monitoring longer-term habitual activity in this age group.