A retrospective study
The study was conducted at a Level III trauma center in Taiwan.
Twenty-five consecutive patients presenting with a Russel-Taylor type 2B fracture were enrolled. All cases were treated with reverse LISS plates. A Modified Radiographic Union Scale for Femur (RUSF), Radiographic parameters, functional scores, and complications were assessed.
Union occurred in 21 cases at an average of 18.8 weeks. The average immediate postoperative neck-shaft angle was 130° (range: 122–135°) compared with 139° (range: 135–141°, p = 0.05) on the contralateral side. Two cases had complications of proximal screws cutting out and two cases had broken implants. Finally, all 4 cases required repeated surgeries (16%). Malunion occurred in 4 patients and early mechanical failure (proximal screws cut out) occurred in 2. There was a significant difference in the purchase index of the proximal screws between cases with redisplacement and those without (26.4 mm and 98.6 mm, p = 0.01).
The use of reverse LISS plate appeared to be an alternative procedure for the specific pattern in the present study. We recommend using this reverse locking plate to treat unstable proximal femoral fractures with meticulous techniques of placing plates. Adequate purchase of the proximal locking screws might decrease the risks of complications.