刊名:Spectrochimica Acta Part B: Atomic Spectroscopy
出版年:1997
出版时间:December 19, 1997
年:1997
卷:52
期:1
页码:2151-2161
全文大小:780 K
文摘
This work presents a versatile method for simultaneous multielemental analysis using detection with photodiode arrays and multivariate calibration techniques. A multidetection system based on an array of 1 024 photodiodes was built and adapted to a commercial plasma emission scanning spectrometer. Spectral data were acquired at low resolution, allowing simultaneous monitoring of a broad spectral range and resulting in very informative but considerably overlapped spectra. Partial least squares and principal component regressions were employed to minimize overlapping problems. A numerical procedure for window selection was also developed.