刊名:Journal of Pharmaceutical and Biomedical Analysis
出版年:2017
出版时间:5 February 2017
年:2017
卷:134
期:Complete
页码:340-345
全文大小:686 K
卷排序:134
文摘
A sensitive, precise and accurate LC–MS/MS method was validated to quantify to bramycin in plasma and microdialysate samples. A very low injection volume (4 μL) was used and a short retention time (∼6 min) was obtained. Microdialysis probe relative recovery for tobramycin is 30% in vitro and in vivo for a flow rate of 1.0 μL min−1. Tobramycin relative recovery is independent of drug concentration and technique used (dialysis or retrodialysis). Tobramycin has a high penetration in interstitial fluid with similar free lung and plasma concentrations in healthy rats.