Pericardium matrix buttressing hinders the stapled bronchial stump healing
详细信息    查看全文
文摘
Biomechanical and histological properties of stapled bronchi with and without bovine pericardial matrix plus collagen reinforcement are analyzed.

Materials and methods

Pneumonectomy with mechanical bronchial suture was performed in the swine model. Pigs were randomly assigned to three groups: three-row staplers without reinforcement (Traditional), with reinforcement (Buttressed) and control “wild type”, non-resected normal bronchus (Normal). Intraoperative test was carried for air leaks at 20/30/40 mm Hg endobronchial pressure. After 60 d, tracheobronchial specimen was harvested, stocked, and analyzed. Tensile test was performed using INSTRON 5965 loading frame machine. Maximal strain resistance value, length of elongation at rupture parameter, and stiffness coefficient (K) were evaluated. Histological analysis was performed. Sample size calculation was assessed (four per group), and the Student t-test was used to statistically evaluate differences in biomechanical variables.

Results

No fistula occurred. Biomechanical analysis showed that maximal strain resistance is 41.22 ± 2.11 N (Traditional), 24.53 ± 3.47 N (Buttressed), and 30.91 ± 0.29 N (Normal); elongation at rupture is 16.01 ± 1.82 mm (Traditional), 12.89 ± 0.48 mm (Buttressed), and 9.32 ± 0.11 mm (Normal). Finally, K is 2.59 ± 0.16 N/mm (Traditional), 1.91 ± 0.33 N/mm (Buttressed), and 3.32 ± 0.01 N/mm (Normal). Stumps without reinforcement proved higher resistance and length of elongation than reinforced ones. Normal bronchial tissue shows the highest stiffness coefficient. Statistical analysis produced significant values for each biomechanical feature. Group Buttressed stumps show greater thickness and a substantial inflammatory reaction with granulation tissue along the whole scar and around areas of discontinuity within the scar, not yet healed.

Conclusions

Reinforcing the mechanical suture line of bronchial stump with bovine pericardial matrix plus collagen shows suboptimal biomechanical and histological characteristics compared to using the stapler alone.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700