A dosimetric evaluation of using a single treatment plan for multiple treatment fractions within a given applicator insertion in gynecologic brachytherapy
详细信息    查看全文
文摘

Purpose

To evaluate the dosimetric impact of using one treatment plan for multiple fractions from a single tandem and ring applicator insertion of high-dose-rate brachytherapy for cervical cancer.

Methods and Materials

Thirteen cervical cancer patients undergoing high-dose-rate brachytherapy were followed. Patients received the total dose from a single applicator insertion in two fractions, given with at least 6 hours apart within 24 hours. The treatment plan was based on a CT scan taken before the first treatment fraction. A second CT was obtained before the second treatment fraction. The co-registered image series were used to evaluate the dosimetric impact of using a single treatment plan for both fractions. Applicator and catheters were measured to quantify interfraction displacement.

Results

When the Day 1 plan was applied to the Day 2 images, high-risk clinical target volume (HR-CTV) coverage was reduced by as much as 17.4 percentage points. The mean decrease was 9.4?¡À?5.0 percentage points (p?<?0.0001). The rectum V75 increase was significant (p?=?0.03), whereas the bladder V75 increase was not significant (p?=?0.28). Volume changes in the HR-CTV contour from Day 1 to Day 2 were also observed (p?=?0.29). Maximum applicator and catheter displacements of 10-30 mm were seen, from Day 1 to Day?2.

Conclusions

When the Day 1 plan was used on the Day 2, the HR-CTV coverage decreased significantly (p?<?0.0001). Our study establishes the need for institutions to evaluate the necessity for replanning based on imaging obtained before each treatment fraction for their gynecologic brachytherapy techniques.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700