We prospectively studied 28 patients implanted with the HeartWare (HeartWare International, Framingham, MA) cfLVAD for up to 360 days. We performed bleeding and coagulation assays 8 times from pre-implant to Day 360 (D360) post-implant, including platelet aggregometry, VWF collagen binding activity-to-antigen (CBA/Ag) ratio, thromboelastography, soluble P-selectin, platelet-specific marker soluble glycoprotein VI (sGPVI), and platelet microparticles. Aortic valve opening was assessed by echocardiography at each assessment. Bleeding and thromboembolic events were documented.
Bleeding events occurred in 14 patients (50%). Maximal platelet inhibition occurred by D30. VWF profile impairment (VWF CBA/Ag < 0.8) was demonstrated in 89% of patients at D30, with subsequent recovery but further deterioration after D180. Bleeding was associated with elevated pre-implant sGPVI (p = 0.008). Pulsatility was associated with higher VWF CBA/Ag (p = 0.02) and a trend to less bleeding.
Third-generation cfLVADs were associated with longitudinal changes in hemostatic markers, and bleeding was associated with elevated pre-implant plasma sGPVI. Further, pulsatility may contribute to recovery of the VWF profile and potentially lower bleeding risk.