Stapled coronary anastomosis with minimal intraluminal artifact: The S2 Anastomotic System in the off-pump porcine model
详细信息    查看全文
文摘
A reliable, easy-to-use, 1-shot anastomotic device will significantly push the barrier for less invasive coronary bypass surgery. The current study was designed to test the safety, efficacy, and early patency of a novel distal anastomotic device.

Methods

The S2 Anastomotic System (iiTech BV, Amsterdam, The Netherlands) was used in 10 consecutive pigs (73 kg) on a mild antiplatelet regimen. In each animal, the device was used to create an internal thoracic artery to left anterior descending bypass on the beating heart. The anastomoses were evaluated intraoperatively (n = 10), at 2 days (n = 2), and at 5 weeks (n = 8) by functional flow measurements, postmortem angiography, and histomorphologic examination.

Results

In all pigs, the S2Anastomic System rapidly created successful anastomoses at the first attempt (graft loading and coronary ischemia time: 1.2 ± 0.3 minutes and 3.0 ± 0.6 minutes) on target vessels of 1.6 to 2 mm inner diameter. There were no technical failures or anastomotic leaks requiring additional sutures. Both intraoperatively and at the time of death, ischemically induced peak hyperemic flow responses demonstrated widely patent bypasses, which were confirmed by postmortem angiography (FitzGibbon grade A, N = 10) and macroscopic evaluation (anastomotic orifice: 2 mm). Histomorphologic evaluation showed a normal healing response with negligible neointima covering the connector and limited streamlining repair tissue formation between the staple-like elements of the connector.

Conclusions

The S2 Anastomotic System consistently created automated, fast, and reliable internal thoracic to coronary artery anastomoses on the porcine beating heart with excellent graft patency and healing characteristics at the 5-week follow-up.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700