Development of a radioxenon measurement system and its application in monitoring Fukushima nuclear accident
详细信息    查看全文
文摘
An HPGe gamma-ray spectrometry system for radioxenon isotopes detection has been developed. The system was comprised of two low-energy planar HPGe detectors that were horizontally and oppositely placed, and a gas source cell made of carbon fiber in both sides was used. The spectra obtained simultaneously by the two HPGe detectors were analyzed by summing spectrum method. After the radioactive concentration of 133Xe was determined with three internal gas proportional counters, the efficiency of 81 keV gamma-ray for 133Xe was calibrated to be 0.50 (1), and the Minimum Detectable Activity (MDA) for 133Xe detection was determined to be 10 mBq during 1 day measurement time. Finally, the system was applied in atmospheric 133Xe monitoring at Xi'an of China after the Fukushima nuclear accident.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700